Dräger · Model Oxylog 2000 Plus
• Built for demanding situations - In the field and in emergency situations, reliability, rugged design and simple operation are vital factors for success. The Dräger Oxylog® 2000 plus was designed to provide the caregiver with powerful and flexible ventilation capabilities under the most demanding conditions. • Intuitive user interface - The large display gives you clear and highly visible information on clinical values, airway pressure and parameter settings, allowing rapid evaluation of the patients condition during hectic situations. Intuitive controls enable operation within seconds after starting the device. • Standardised equipment - The Oxylog® 2000 plus uses the same standardised reusable and disposable hoses as the Oxylog® 3000, letting you simplify your inventory and workflow. • Ventilation essentials at your fingertips - The Oxylog® 2000 plus gives you not only a selection of volume-controlled modes but also offers support modes for both invasive and non-invasive ventilation. Pressure Support and Non-Invasive Ventilation are available, enabling support for patients with insufficient breathing and to help prevent intubation at the earliest stage possible. • Rugged, compact and highly portable - Easy to carry and simple to mount, the Oxylog® 2000 plus was specifically designed to be taken nearly anywhere. Built from extremely durable, high-impact materials, the unit can stand up to even the most challenging environments. The internal battery provides up to 4 hours of autonomous transport capability.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
Consumable, replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable, replace per PM schedule. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.