Dräger · Model Infinity CNAP SmartPod
Integrates light-emitting diode & detector into sensor cuff. Blood pressure can then be calculated beat-to-beat, after calibration with the built-in standard oscillometric measurement (NBP) of the monitor – and a high resolution, real-time arterial pressure waveform is displayed. Bridges the gap between NBP and arterial line. The CNAP solution provides the clinical utility (i.e., beat-to-beat vigilance) of the invasive arterial line – with user convenience and simplicity of a noninvasive cuff. It bridges the gap in many situations when NBP is not enough, and an arterial line is just too invasive and complicated.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: IBP / PiCCO / CNAP trace offset, "zero required" persists, values implausible.
Root cause: Transducer not open to atmosphere during zero, a damaged cable, or a module input fault.
Symptoms: Host monitor does not display the module or pod.
Root cause: Docking / pod connector or cable fault.
Symptoms: Damped, noisy or flat trace.
Root cause: Air / clot in the line, a loose connection, or a cable fault.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set | Frequent wear item | $$ | Easy |
| Pressure transducer set | Consumable | $ | Easy |
| Module input board | Major service part | $$$ | Difficult |
| Module / pod docking connector | Major service part | $$ | Moderate |
| Reusable IBP cable | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.