DeVilbiss · Model 5 Liter Oxygen Concentrator
Made in the USA of US and Imported Parts. Exclusive DeVilbiss OSD® (Oxygen Sensing Device) increases accuracy and reliability for longer service intervals. Noticeably quiet operation – 15% typical sound quality improvement. Readily accessible patient controls and protected cannula fitting. Recessed humidifier nook prevents damage. Easy to assemble/disassemble two-piece cabinet for ease of maintenance. Lockable flow meter - Pediatric flow meter available. Convenient top and side handles for easy transport. Accommodates up to a 50-foot cannula (15.3 meters).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Oxygen-concentration alarm, delivered O2 below spec at flow.
Root cause: Aged sieve beds (molecular sieve), a compressor / valve fault, an air leak, or an inlet-filter blockage.
Symptoms: Weak or no flow, compressor noisy or hot, thermal cut-out.
Root cause: Worn compressor / valve heads, or a cooling-fan fault.
Symptoms: Flow will not adjust or does not match the setting.
Root cause: Worn flow-control valve or flowmeter.
Symptoms: Reduced performance, compressor working hard.
Root cause: Dirty gross-particle / cabinet / final bacteria filter.
Symptoms: Loud operation, rattling.
Root cause: Worn compressor mounts or a loose panel.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Molecular sieve beds | Periodic replacement | $$$ | Difficult |
| Compressor / valve-head rebuild kit | Periodic replacement | $$$ | Difficult |
| Inlet / cabinet / final filters | Frequent wear item | $ | Easy |
| Flow-control valve / flowmeter | Major service part | $$ | Moderate |
| Cooling fan | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.
The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.
An unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patient gas pathway)