FC1Battery near depletion / Replace with charged battery or switch to alternate power source.
Battery near depletion / Replace with charged battery or switch to alternate power source.
FC2Battery near depletion / Replace with charged battery or switch to alternate power source.
Replace with charged battery or switch to alternate power source.
FC3Battery Depleted / Replace with charged battery or switch to alternate power source
Replace with charged battery or switch to alternate power source
FC4TOC does not recognize installed battery / Verify battery is fully inserted into the TOC. Remove and reinstall battery. Verify battery is for use with the PM4400 TOC.
Verify battery is fully inserted into the TOC. Remove and reinstall battery. Verify battery is for use with the PM4400 TOC. Replace battery with new battery.
FC51. Allow battery to warm up and charging will start automatically after alarm is acknowledged. 2. Replace with another battery. 3. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Allow battery to warm up and charging will start automatically after alarm is acknowledged. Replace with another battery. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC61. Allow battery to cool, charging will start automatically after alarm is acknowledged. 2. Replace with another Battery. 3. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Allow battery to cool, charging will start automatically after alarm is acknowledged. Replace with another Battery. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC7TOC or parts of TOC are overheating / Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC8TOC or parts of TOC are overheating / Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC9TOC or parts of TOC are overheating / Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC10TOC or parts of TOC are overheating. TOC will shutdown if condition cannot be resolved./ Power off TOC and allow to cool 5 to 10 minutes . Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool 5 to 10 minutes. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC11TOC or parts of TOC are overheating. TOC will shutdown if condition cannot be resolved./ Power off TOC and allow to cool 5 to 10 minutes . Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool 5 to 10 minutes. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC12TOC or parts of TOC are overheating. TOC will shutdown if condition cannot be resolved./ Power off TOC and allow to cool 5 to 10 minutes . Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
Power off TOC and allow to cool 5 to 10 minutes. Ensure the TOC is well ventilated. Return the TOC to your prescribed dose. If alarm condition persists, contact your Healthcare Provider. Switch to an alternate oxygen source.
FC13TOC is unable to detect patient’s breath with TOC in Pulse mode / Verify cannula length is a maximum of 7 feet. Verify cannula connection is tight. Verify cannula nasal prongs are properly placed into nostrils. Breathe through the nose and not the mouth. Use “nasal cannula” only; do not use cannula with mask.
Verify cannula length is a maximum of 7 feet. Verify cannula connection is tight. Verify cannula nasal prongs are properly placed into nostrils. Breathe through the nose and not the mouth. Use “nasal cannula” only; do not use cannula with mask.
FC141. Reduce level of activity to lower breath rate. 2. If alarm persists, change to alternate oxygen source and contact your Equipment Provider.
Reduce level of activity to lower breath rate. If alarm persists, change to alternate oxygen source and contact your Equipment Provider.
FC15Cannula blocked or pinched / Verify the cannula is not blocked or pinched. Replace cannula if damaged.
Verify the cannula is not blocked or pinched. Replace cannula if damaged.
FC16TOC does not recognize the position of the Settings Dial / Ensure Settings Dial pointer is not between settings. Verify dial is pointing directly at setting.
Ensure Settings Dial pointer is not between settings. Verify dial is pointing directly at setting. If alarm condition persists, select a different setting and then return to prescribed setting.
FC17Oxygen Concentration is below 83% / Continue to use TOC and contact your Healthcare Provider. Switch to an alternate oxygen source.
Continue to use TOC and contact your Healthcare Provider. Switch to an alternate oxygen source.