Datex Ohmeda S5 ADU Carestation The is an innovative approach to anesthesia delivery, which utilizes electronic control for both gas delivery and vaporization. This enables full benefit of the automatic data capture of care information management and minimizes manual documentation. Data integration allows you to configure the data on full color screens – anesthesia delivery settings on one display and patient monitored values on the other. An additional display is available for integrated anesthesia record. The ADU Carestation can be linked, through the Datex-Ohmeda Anesthesia Monitor, to a network of other carestations and allows interface with the hospital-wide information systems. Multiple product options and software-driven technology helps enable future upgrades of the system to suit your preferences. Integration of anesthesia delivery.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace on schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
| CO2 absorber canister Consumable, replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.