Intuitive and highly configurable touch screen Graphs may be scaled and enlarged to full screen to improve visibility. Using the pause button, you can pause the screen and review the last 60 seconds of patient data. In situations where you may not be directly in front of the screen, or during teaching situations where there are several people in the room, the large-font screen can provide greater visibility to patient data Screen configurability Several configuration options enable you to see as little or as much information as you need. Patient data in the top banner may be configured to the patient need or to match your workflow document. Additional patient data can be accessed by scrolling right to left in the upper right corner. Patient data cells may be left blank to reduce the amount of visible patient data.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable - replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical ventilation due to a software error.