CAS Medical · Model 740
Fast, accurate blood pressure readings for neonates through adults even in the most challenging situations such as high motion or low perfusion. Monitor SpO2 and temperature either continuously or in "spot-check" mode. Accurate oral temperature in just 4 seconds. Ideal for medical/surgical patients on pain medications including PCA pumps, for hospital or office-based endoscopy procedures, and for the EMS applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible or fail against a reference.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer / calibration fault.
Symptoms: No saturation value or a poor pleth with a good site.
Root cause: Worn SpO2 sensor / cable, or the SpO2 module.
Symptoms: No or implausible temperature.
Root cause: Probe or cable fault.
Symptoms: Short battery run time, will not power on or charge.
Root cause: Aged battery pack.
Symptoms: No or faint strip, paper jam.
Root cause: Thermal print head or paper handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| SpO2 sensor / cable | Frequent wear item | $$ | Easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Software: Audible alarm may be silenced if changed from the factory setting
From BioLab, for work done with this kind of equipment.