CareFusion · Model ReVel
Integrating tools that help clinicians make safe predictive decisions with Spontaneous Breathing Trial (SBT), studies have found that patients who tolerate a SBT for 30-120 minutes were found to have successful discontinuation of ventilation at least 77% of the time1. Integrated Pulse Oximeter allows monitoring of patient’s SpO2 and pulse rate during transport or in lieu of central monitoring.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Self-test aborts with a diagnostic code, circuit-leak or compliance test fails, ventilator not released for use.
Root cause: Patient-circuit leak, an exhalation-valve / flow-sensor fault, or an internal pneumatic / electronic fault (see the diagnostic-code table).
Symptoms: FiO2 not matching the setting, low or high supply-pressure alarm, gas-mix alarm.
Root cause: Blender fault, a wall-supply / hose issue, or a failing internal compressor / turbine.
Symptoms: Monitored tidal volume, pressure or flow implausible, frequent nuisance alarms.
Root cause: Flow sensor(s) or the internal pressure transducers aging or contaminated.
Symptoms: PEEP not held, auto-PEEP, high circuit pressure, exhalation-valve alarm.
Root cause: Worn or wet exhalation-valve diaphragm / assembly.
Symptoms: Short battery run time, will not ride through a power interruption, will not charge.
Root cause: Aged internal / external battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly | Frequent wear item | $$ | Fairly easy |
| O2 sensor (galvanic cell) | Frequent wear item | $$ | Fairly easy |
| Internal compressor / turbine | Major service part | $$$$ | Expert-level |
| Battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
CareFusion has identified a potential for damage to the power connector on the EnVe and ReVel ventilators due to misconnection by the operator. The input port on the ventilators must have the adapters correctly aligned to function properly. Incorrect alignment may cause the pins in the connector to bend, causing the PTV AC power adapter and/or PTV Auto Lighter Power Cord to be shortened, and pre
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio