CareFusion · Model Pulmonetics LTV 1100
• High performance ventilation in a small lightweight package (10” x 12” x 3”, 13.4 lbs). • Turbine technology allows the LTV Series ventilator to operate without an external compressed gas source. • CPAP5, SIMV6, Control, Assist / Control and Apnea Backup ventilation modes. • NPPV7 mode ventilation, providing an alarm package suitable for mask ventilation of patients that do not require life support ventilation. • Volume Control, Pressure Control (optional) and Pressure Support ventilation. • Variable alarm settings including High Peak Pressure, Low Peak Pressure, Low Minute Volume, Apnea, High Breath Rate, and High PEEP. • Oxygen Blending from a High-Pressure Oxygen source, Low-Pressure Oxygen Bleed-in, O2 Flush, and O2 Cylinder Duration Monitoring on the LTV 1000, and Low-pressure Oxygen Bleed-in on the LTV 950 and 900. • Lockable front panel controls. • Monitors for Breath Rate (f), I:E Ratio, MAP, Minute Ventilation (VE), PEEP, PIP and Tidal Volume (Vte). • Real-time patient circuit pressure display with Peak Inspiratory Pressure indicator. • Variable termination conditions for Pressure Support breaths, including maximum inspiratory time termination and percentage of peak flow. • Selectable Percentage of Peak Flow termination for Pressure Control breaths. • Leak Compensation to improve triggering when a circuit leak is present. • Single or dual tone output capabilities. • Operation from a variety of power sources including AC power, internal battery and external DC power sources.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.