BPL Medical Technologies · Model Leoni Plus HFO
Suitable for long term ventilation of premature babies and infants, neonates and paediatric patients weighing up to 30 Kg. Ventilation modes include nCPAP, IPPV/IMV, S-IPPV, SIMV and PSV for premature and infants. Assisted Ventilator modes include Volume-Controlled Tidal Volume guarantee. It works by setting a fixed tidal Volume (VT) that the ventilator will deliver at set intervals of time or when the patient initiates a breath with 10 sensitivity levels. Patient proximate leakage is managed with a compensated flow or a volume trigger and is calibrated within seconds of ventilation interruption. Graphic display, graphic trend and log data of spontaneous breaths displayed simultaneously. A 12” colour display and a detachable touchscreen control unit for displaying crucial parameters, graphs, loops etc. It consists of an Oxygen/ compressed air mixer with an integrated oxygen monitor. Integrated rechargeable battery. Upgradeable to Double Membrane Extremely powerful, High frequency Oscillatory ventilation, with time-controlled boosting of average pressure, and adjustable I/E ratio.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
Consumable - replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable - replace per PM schedule. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.