Boyd Industries · Model S3100LC
Lockable swivel base of capable of 90 degrees of rotation Lift column base allows chair to be raised or lowered in a fixed field of view Low profile, double articulating headrest Easy access membrane switches on left and right side of chair.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Base, back or a positioning function does not respond to the foot control or touchpad.
Root cause: Foot control / touchpad fault, a hydraulic pump / valve, an actuator, or a control-board fault.
Symptoms: Base slowly lowers under load after positioning.
Root cause: Internal leak in the hydraulic system or a failing check valve.
Symptoms: Tilting back or the headrest slips under load.
Root cause: Worn lock or friction hardware.
Symptoms: Loud pump on lift, slow rise.
Root cause: Worn pump or low hydraulic fluid.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Hydraulic pump / motor | Major service part | $$$ | Difficult |
| Hydraulic control valve | Major service part | $$ | Difficult |
| Foot control / touchpad | Periodic replacement | $$ | Fairly easy |
| Back / tilt actuator | Major service part | $$$ | Difficult |
| Headrest / back lock hardware | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.