Beijing ADSS Development Co., Ltd. · Model FG 2000-D
Diode Laser Therapy Machine.
Intended use: A mains electricity (AC-powered) device assembly in which input energy is used to excite a diode to emit a high-power therapeutic laser beam designed to perform dermatological procedures such as body hair removal (e.g., armpit, leg/arm, back, chest, bikini line, and face) and/or skin resurfacing (e.g., treatment of benign lesions, wrinkle reduction). It is intended to be used by a trained professional to emit pulsed light of a specific wavelength (e.g., around 808 nm or 1060 nm), delivered by a dedicated handpiece. It typically includes a control unit with dedicated software and user interface, delivery handpiece, and footswitch.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Weak or no output at the handpiece / fibre, treatment ineffective, power not reaching the set value.
Root cause: Diode laser module aging, a damaged delivery fibre / handpiece, or a driver / power-supply fault.
Symptoms: Unit shuts down after a period, over-temperature warning.
Root cause: Chiller / TEC cooling fault, low coolant, or blocked airflow.
Symptoms: Burn marks at the fibre tip, high loss, intermittent output.
Root cause: Damaged or contaminated fibre tip or connector.
Symptoms: Laser will not fire, key / interlock / footswitch error.
Root cause: Footswitch, door interlock, or key-switch fault.
Symptoms: No or dim aiming beam.
Root cause: Aiming-beam diode or its alignment.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Power re-calibration required after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Diode laser module Power re-calibration required after replacement. | Periodic replacement | $$$$ | Expert-level |
| Delivery fibre / handpiece | Frequent wear item | $$ | Fairly easy |
| Cooling (TEC / chiller) components | Periodic replacement | $$$ | Difficult |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Aiming-beam diode | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio