Beckman Coulter · Model Optima L-XP Series
A smarter ultracentrifuge. For smarter results - The advanced Optima eXPert Software provides unprecedented centrifugation capabilities. It assists you in developing the most efficient centrifugation methods, selecting the rotor and labware for a given application, then simulating the results before performing separations of proteins and sub-cellular particles. Password protection for up to three levels of users allows you to grant the right level of access to the right people. With this degree of centrifugation knowledge, you get more efficient use, faster turnarounds and peace-of-mind security. Putting a smart new spin on your research - The innovative rotors Beckman Coulter built for the Optima L-XP are designed, manufactured and tested as a system from the inside out to make sure you get the highest quality of separation in the shortest possible time-safely and reliably. The increased efficiency of our SW 32 Ti series rotors can reduce run times. And thanks to a slight angle that positions contaminants away from the bands of interest, our patented NVT™ rotors deliver the highest purity in the shortest possible time. Innovative drive technology - Optima L-XP utilizes the most advanced materials and technology to achieve the high performance needed - up to 100,000 rpm - for extending the boundaries of research. Plus, the air-cooled, patented design of our drives delivers the most energy-efficient ultracentrifugation system ever made. And the imbalance-tolerant Optima L-XP drive safely accommodates tubes that are under- or over-filled by as much as 10%. The best place for your samples - Since they're specially designed for innovative application solutions, our patented tubes deliver improved efficiencies, including Quick-Seal® for biocontainment, g-Max™ for the greatest volume efficiency, konical™ seal for the most concentrated pellet and OptiSeal™ for the easiest sealing. And the design of our one-touch OptiSeal tubes delivers fast, reliable sealing--just insert the plug and press into place, and the centrifugation process itself provides guaranteed protection for your samples.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Vacuum-fault alarm, long pump-down, run aborts.
Root cause: Vacuum-pump wear (low oil / worn vanes), a chamber / feed-through seal leak, or a valve fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Vacuum pump oil & O-ring kit | Periodic replacement | $$ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio