EPPENDORF · Model Centrifuge 5810 / 5810 R
Interchangeable swingbucket and high-speed fixedangle rotors. Spin up to 14,000 rpm, or 20,800 x g. Low-profile 11.2 in/28 cm access height for easy sample loading. Automatic rotor recognition senses rotor type to set maximum allowable speed - ensures safety. Fast, simple programming of time, temperature (refrigerated models), speed/ g-force and radius value. 34 user-defined programs. Adjustable-speed Short Spin button. Ten acceleration and braking rates. Simple operation. Speed, g-force and radius correction values can be entered. Values can be changed during centrifugation. Temperature range from –9 °C to 40°C, with Fast Temp function for rapid cooling of the centrifuge chamber. Standby cooling maintains temperature when not in use. Speed can be set from 200 rpm to maximum speed in increments of 10 rpm. Short Spin button. Selectable program memory for up to 34 individual user programs. Ten acceleration and deceleration rates for sensitive sample material.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Temperature high during a run, or slow to pre-cool.
Root cause: Compressor, condenser fan, low refrigerant, or a temperature-sensor fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Condenser fan / refrigeration components | Major service part | $$$ | Difficult |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.