Beckman Coulter · Model Epics XL and XL-MCL
3-color fluorescence with 4-color option Air-cooled 488nm 15mw laser 10u x 90u light beam Detector Range: 300nm-800nm Flow-Cell Size: 250u x 250u BioSense Max analysis rate: >3300 cells/sec Min Cell Size, um: 3300 cells/sec Color dot density; contour, isometric, tomogram, overlay, legend and projection plots. PRISM displays; color precedence or color-blend gating; automatic Boolean logic gate creation; 32 gates per file, real-time statistics. Intel Pentium processor and 4 network transputers for digital signal processing. Signal-intensity display, biohazard-safe triple-fluid (sheath/cleanse/waste) system with automated functions and variable flow rate; auto setup; SYSTEM II software with acquisition, list mode, multigraph, reports and utilities. Features FastSet and CellTracker. Opt tetraONE SYSTEM software automates 4-color acquisition and analysis. Retic-ONE SYSTEM software for reticulocyte enumeration. The XL system features the capability to analyze up to 4 colors of immunofluorescence from a single air cooled laser. The instrument is self contained and biohazard safe. The user retains the flexibility to change filter elements for versatility in research settings. For high volume users, the XL-MCL system offers walk away sample handling for superior efficiency with the Multi Carousel Loader (MCL). The XL remains the only flow cytometer to offer state-of-the-art Digital Signal Processing (DSP) for reliable linearity and drift-free amplification and compensation. The XL SYSTEM II™ software fully automates instrument set-up and compensation for 2, 3 and even 4-color applications. The XL is available in 3 or 4-color configurations, with or without the MCL. Ask your Beckman Coulter representative about the configuration that best meets your laboratory's needs.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Bead CVs out of spec, scatter or fluorescence signal low or noisy, compensation drift.
Root cause: Flow-cell contamination or an air bubble, laser degradation / misalignment, or an optics / PMT fault.
Symptoms: Unstable event rate, boost / prime failures, sheath or waste alarm, clogs.
Root cause: Sheath / waste tank issue, a peristaltic / vacuum fault, or a partially blocked sample line / flow cell.
Symptoms: No or low event rate, probe-position error, carryover.
Root cause: Clogged or bent sample probe, or a clot.
Symptoms: A laser-line-specific parameter fails, laser-error message.
Root cause: Laser at end of life or a power-supply / interlock fault.
Symptoms: Acquisition hangs, data will not export, instrument not recognised.
Root cause: Workstation PC or the acquisition software / link.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow cell | Periodic replacement | $$$ | Difficult |
| Sample probe / SIP | Periodic replacement | $$ | Moderate |
| Peristaltic pump tubing | Frequent wear item | $ | Moderate |
| Sheath / waste filters | Frequent wear item | $ | Easy |
| Laser | Periodic replacement | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The firm has confirmed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. This issue is limited to only protocols utilizing the Ratio parameter.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.