Beckman Coulter · Model Epics XL and XL-MCL
3-color fluorescence with 4-color option Air-cooled 488nm 15mw laser 10u x 90u light beam Detector Range: 300nm-800nm Flow-Cell Size: 250u x 250u BioSense Max analysis rate: >3300 cells/sec Min Cell Size, um: 3300 cells/sec Color dot density; contour, isometric, tomogram, overlay, legend and projection plots. PRISM displays; color precedence or color-blend gating; automatic Boolean logic gate creation; 32 gates per file, real-time statistics. Intel Pentium processor and 4 network transputers for digital signal processing. Signal-intensity display, biohazard-safe triple-fluid (sheath/cleanse/waste) system with automated functions and variable flow rate; auto setup; SYSTEM II software with acquisition, list mode, multigraph, reports and utilities. Features FastSet and CellTracker. Opt tetraONE SYSTEM software automates 4-color acquisition and analysis. Retic-ONE SYSTEM software for reticulocyte enumeration. The XL system features the capability to analyze up to 4 colors of immunofluorescence from a single air cooled laser. The instrument is self contained and biohazard safe. The user retains the flexibility to change filter elements for versatility in research settings. For high volume users, the XL-MCL system offers walk away sample handling for superior efficiency with the Multi Carousel Loader (MCL). The XL remains the only flow cytometer to offer state-of-the-art Digital Signal Processing (DSP) for reliable linearity and drift-free amplification and compensation. The XL SYSTEM II™ software fully automates instrument set-up and compensation for 2, 3 and even 4-color applications. The XL is available in 3 or 4-color configurations, with or without the MCL. Ask your Beckman Coulter representative about the configuration that best meets your laboratory's needs.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
Consumable, replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
| Peristaltic pump tubing Consumable, replace proactively. | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The firm has confirmed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. This issue is limited to only protocols utilizing the Ratio parameter.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →