Beckman Coulter · Model Allegra 6 Series
Ideal for Cell Culture, Plasma and Other General-Purpose Separations. Exclusive ARIES Smart Balance Rotor recognizes and corrects rotor imbalance up to 50 grams per bucket set. Process up to 56 x 15 mL conical tubes or 12 standard microplates. Spins tubes, bottles, microplates, specialty bottles and blood bags. Interchangeable rotors available. Performing from 2,010 to 5,710 x g. Offered in refrigerated or non-refrigerated benchtop and refrigerated knee-well models.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Chamber will not reach set temperature.
Root cause: Compressor, condenser, or a sealed-system fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still in progress.
Symptoms: RCF / RPM or chamber temperature does not match the setting.
Root cause: Tacho / speed-sensor or temperature-sensor drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor use; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed / temperature sensor | Major service part | $$ | Moderate |
| Refrigeration components | Major service part | $$$$ | Expert-level |
| Rotor / rotor O-rings Log rotor use; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio