Beckman Coulter · Model ACL Futura
A high functioning coagulation analyzer The fully automated instrument assures you better performance in screening assays, functional and chromogenic assays, and specific factor assays. IL ACL Futura’s compact size combined with its enhanced features and capabilities such as random access, continuous sample loading, true walkaway productivity, and flexibility makes it ideally suited for use in any clinical laboratory setting. It has the ability to store up to 120 test samples in primary sampling tubes or sample cups. This system has automated positive-displacement pipetting for providing reliable test results. All dilutions for calibration or sample analysis are done automatically by the system itself. For high-throughput sampling, the instrument makes use of two pipettes. Each pipettor comes with a liquid sensor in order to adjust the system operation to meet the different liquid levels, sample sizes and containers. This instrument has 34 refrigerated reagent positions and the reagents in their own vials can be kept directly on the instrument. This instrument has state-of-the-art PC driven software which makes it appropriate for any modern lab setting. Sample analysis order and two reading stations enable conducting chromogenic and clotting analysis. A built-in quality control program with database and archiving capability along with Levey-Jennings charts are also included in the device.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.