Sysmex · Model CA 500
A compact fully-automated instrument capable of 5-parameter random analysis for In Vitro Diagnostic use. This instrument incorporates the latest technologies as represented by micro-computers, thus enabling analysis of multiple parameters with increased flexibility. Of PT, APTT, Fbg, Thrombin Time (Coagulation Method), and Antithrombin III (Chromogenic Method: Can be analyzed only with CA-530, CA-540, CA-550 and CA-560), D-Dimer (Immunology Method: Can be analyzed only with CA-550 and CA-560), etc., this instrument is able to analyze 5 parameters simultaneously. In addition, it has a number of functions including preferential processing of STAT samples and a built-in quality control function. Moreover, it allows analyzed data to be displayed and printed out together with reaction curves, thus making it possible to obtain highly reliable analysis results.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.