Baxter International · Model IQ Infusion System
Built-in DeviceVue Asset Tracking Application, which displays pump status and location data on a PC, tablet or smartphone without the need to interface with third-party real-time location systems Spectrum IQ Infusion Dashboard with Charge Capture, which automates the capture of start, stop and duration times, helping to optimize outpatient IV billing Alarm and alert routing designed to help reduce “alert fatigue” among nursing staff by sending alarm start and stop messages directly from the pump bedside to secondary Alarms Management Systems, including smartphones or work stations Enhanced data analytics and near real-time infusion data accessible from a single, centralized screen Single set technology that can help yield up to 53 percent cost savings in IV tubing sets and up to 30 percent reduction in IV tubing usage.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The pump alarms and displays an error message / code and stops.
Root cause: A subsystem fault indexed in the service manual - motor / drive, a sensor, set-detection, or a board.
Symptoms: Repeated occlusion alarms, delivery stops, or an occlusion is missed.
Root cause: Kinked or clamped line, a closed roller clamp, a positional patient tube, pressure-sensor drift, or an improperly loaded set.
Symptoms: Frequent air-in-line alarms, delivery stops.
Root cause: Air in the set, an improperly loaded set, or an ultrasonic air-detector fault.
Symptoms: Delivered volume outside tolerance on the annual test, especially at low rates.
Root cause: Worn pumping mechanism / plunger drive.
Symptoms: Short run time, low-battery early after a full charge.
Root cause: Aged battery pack.
Symptoms: Set will not load, door will not latch, plunger grip poor.
Root cause: Worn door latch, plunger jaws, or anti-free-flow mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Pumping mechanism / rotor / plunger drive | Major service part | $$$ | Difficult |
| Pressure / air-detector sensor assembly | Major service part | $$ | Moderate |
| Door / latch / anti-free-flow assembly | Major service part | $$ | Moderate |
| Drive motor / gearbox | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Pumps were released without full testing being performed, including occlusion alarm testing.
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Certain Spectrum infusion pumps may have an incorrect version of software.
There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to cleaning, or as a result of an IV fluid spill.
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
Devices were identified as released after repair without full testing being performed, which includes flow testing.
The door on the Spectrum IQ Infusion pump may not be able to fully close.
The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system.
Improperly performed testing prior to release
There is a potential for cracks on the mount of the front panel of the device.
One device was improperly performed testing prior to release from a Service Center.
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Potentially defective component in the AC power adapter which may fail over time.
Battery performance issues related to prolonged storage.
There is a potential software error during programming.
Baxter is updating the device IFU for safety communication regarding cleaning practices.
If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
The flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not identify issues with pumps that could cause inaccuracies in flow rates which, in turn, could lead to over- or under- infusion. Potential risk to the patient depends on a number of factors including treatment and patient comorbidities. Depending on these factors, the patient could experience serious adverse health consequences.
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.
Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.
Baxter Healthcare Corporation has issued an Urgent Device Correction for the SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library due to repeated System Error 322 occurrences.
Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality. Complaint rates for ruptures have since decreased. Baxter wants to ensure that no affected product remains in the field.
Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.
Some of the Single Day Infusor System 2 mL/h have a foil over-pouch that was incorrectly labeled as the Half Day Infusor System 5 mL/h, product code 2C1073KJP, lot 10H054.
The FloGard Infusion Pump was released from service without having the Current Leakage test performed. This could lead to the patient/user to receive an electrical shock.
If the manifolds are connected in series, there may be a risk of the manifolds loosening or disconnecting at the luer connection between them.
During packaging testing, Baxter observed tears and pinholes in Chevron-style pouches for certain infusion disposable products. The pinholes and tears could compromise the sterile barrier properties of the pouch.
There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage
Leaking: leaks at the tubing flow restrictor connection to either the coupler or the male luer during filling.
Delivery of unrequested PCA doses of medication due to damaged PCA cords, partial sticking of PCA button, and fluid ingress into the pump or PCA button.
When infusing at flow rates below 0.5 mL/hr, the upstream occlusion alarm of the pumps may not detect the occlusion.
Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
There is an overinfusion condition in the Intermittent mode where the pump may continue to deliver at the higher dose rate instead of switching to the lower keep open rate once the first dose is completed.
Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.
Premature motor failure.
From BioLab, for work done with this kind of equipment.