Bard · Model Finesse Ultra
Acquire multiple samples rapidly - sampling cycle time is just 10 seconds Less time, less trauma, and less anxiety for an improved patient experience.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No keypad lights after loading a probe, or lights do not turn solid green.
Root cause: Probe cover or locking tab not fully seated, a reused probe, insufficient driver charge, or a probe fault.
Symptoms: Battery indicator does not rise on the charging stand.
Root cause: Charging-stand contacts, the stand power supply, or an aged driver battery.
Symptoms: No lights, no response - driver idle more than 120 seconds.
Root cause: Sleep mode - move the driver or place it on the charging stand; if still dead, battery or electronics.
Symptoms: Prime/pierce or sample does not complete.
Root cause: Drive motor or a damaged probe.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Bard Service and Repair for internal work.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable driver battery Bard Service and Repair for internal work. | Periodic replacement | $$ | Moderate |
| Charging stand | Major service part | $$ | Easy |
| Single-use probe | Consumable | $ | Easy |
| Drive motor assembly | Major service part | $$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.