Hologic · Model Trident
Superb image quality for rapid verification with maximum confidence. One-touch x-ray control with AEC for fast image acquisition. Large, 12 cm x 14 cm active imaging area; accommodates a wide range of surgical specimens and core biopsy samples. Innovative Enhanced Visualization tool with five levels of image optimization for added sharpness and lesion conspicuity. Fully integrated, maneuverable and ergonomic workstation. User-friendly operator interface with minimal procedure steps to streamline workflow. Ideal for radiologists, breast surgeons and pathologists.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blank, noisy or low-contrast specimen image, exposure inhibited.
Root cause: X-ray source aging, a flat-panel detector fault, or a generator / interlock fault.
Symptoms: Fixed-pattern artefacts, non-uniform image, calibration overdue or failing.
Root cause: Detector gain / offset calibration or a bad pixel map.
Symptoms: Drawer will not open / close, door-interlock error, exposure blocked.
Root cause: Drawer mechanism or a door / interlock switch.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| X-ray source | Major service part | $$$$ | Expert-level |
| Flat-panel detector | Major service part | $$$$ | Expert-level |
| Drawer / door interlock switch | Major service part | $$ | Moderate |
| Workstation PC | Major service part | $$$ | Moderate |
| Generator components | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.