Tissue acquisition occurs every 4.5 seconds, typically lasts less than 30 seconds total. Easily delivers local anesthetic without interruption. Closed system reduces contamination risk. Fully disposable device and tubing minimizes staff and patient exposure to biohazard Simple one minute set-up and clean-up. No software required for user to program or operate the console.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: System fails to achieve or maintain required vacuum pressure; biopsy cycle aborts or completes with insufficient tissue sample
Root cause: Leak in disposable tubing set, clogged filter in air motor regulator, or relief regulator pressure out of specified 70-72 PSI range
Symptoms: Compressor runs continuously; audible grinding or rattling from air motor housing; vacuum pressure drops below operating threshold
Root cause: Worn internal vanes in air motor, debris ingestion from compromised filter, or inadequate ventilation of console enclosure
Symptoms: Biopsy trigger does not activate vacuum; requires multiple presses or specific angle to function
Root cause: Footswitch internal contact wear, disconnected ribbon cable at console rear, or fluid ingress from inadequate sealing
Symptoms: Vacuum pressure reads outside 70-72 PSI range during operation; system over-pressurizes or under-performs
Root cause: Regulator knob loosening from vibration, diaphragm fatigue, or contamination in pressure sensing line
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Per-procedure consumable; inspect for kinks or cracks before each use
Located on left side of air motor regulator; replace if pressure checks fail cleaning
Adjustable between 70-72 PSI per section 3.4.1; older units may have right-side mounting with blue cap
Check ribbon cable connection at console rear before assembly replacement
Standard IEC inlet; replace if continuity fails or insulation degraded
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable tubing set (vacuum line and probe connector) Per-procedure consumable; inspect for kinks or cracks before each use | Consumable | $ | Easy |
| Air motor filter element Located on left side of air motor regulator; replace if pressure checks fail cleaning | Frequent wear item | $ | Fairly easy |
| Relief regulator assembly Adjustable between 70-72 PSI per section 3.4.1; older units may have right-side mounting with blue cap | Periodic replacement | $$ | Moderate |
| Footswitch assembly Check ribbon cable connection at console rear before assembly replacement | Periodic replacement | $ | Fairly easy |
| Power cord and inlet filter Standard IEC inlet; replace if continuity fails or insulation degraded | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.