B. Braun · Model Outlook 200
Bedside barcode verification Promotes safer administration practices by reducing manual programming steps using barcoding with patient matching Promotes integrated safety by assigning dose limits to scanned drugs Promotes easy titration by prompting for the starting dose on “titrate” orders Visually indicates dose selection with DoseGuard™ software Easy to learn Manual operation RATE, VOLUME, RUN Automatic Dose Rate Calculator Large, highly visible keys Reduce nuisance air-in-line alarms with air-trapping cassette Stackable modular design reduces bedside cord clutter and errors due to channel confusion Comprehensive line of needle-free administration sets include UV light-resistant tubing that protects therapy efficacy.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The pump alarms and displays an error message / code and stops delivery.
Root cause: A subsystem fault indexed in the service manual - motor / drive, a sensor, set-detection, or a board.
Symptoms: Repeated occlusion alarms, delivery stops, or an occlusion is missed.
Root cause: Kinked or clamped line, a closed roller clamp, a positional patient IV, a pressure-sensor drift, or an improperly loaded set.
Symptoms: Frequent air-in-line alarms, delivery stops.
Root cause: Air in the set, an improperly loaded set, or an ultrasonic air-detector fault.
Symptoms: Delivered volume outside tolerance on the annual test, especially at low rates.
Root cause: Worn pumping mechanism / plunger drive, or (syringe pumps) a worn plunger grip / drive train.
Symptoms: Short run time, low-battery early.
Root cause: Aged battery pack.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Pumping mechanism / door assembly | Major service part | $$$ | Difficult |
| Pressure / air-detector sensor assembly | Major service part | $$ | Moderate |
| Drive train / motor | Major service part | $$$ | Difficult |
| Administration / syringe set | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.