ASP · Model Sterrad 100S
Endorsed by leading Medical Device Manufacturers on their IFUs. Sterilizes a wide range of instruments, including da Vinci® 3-D endoscopes*. One button operation makes it easy to use.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle cancels with a fault code, load not sterile, must be re-run.
Root cause: Insufficient vacuum, a load / packaging that absorbs peroxide, a peroxide-cassette / injection fault, or a plasma-generation (RF electrode) fault.
Symptoms: Vacuum-fault alarm, long pump-down, cycle aborts.
Root cause: Vacuum-pump wear (low oil / worn vanes on older D16 pumps), a chamber-door seal leak, or a valve fault.
Symptoms: Low-concentration or monitor alarm, cassette not read.
Root cause: Expired or empty peroxide cassette, a cassette-reader fault, or a peroxide-monitor sensor.
Symptoms: Peroxide not vaporised or injected correctly, cycle fault.
Root cause: Vaporizer plate contamination or a heater / valve fault.
Symptoms: No cycle printout, unresponsive display.
Root cause: Printer or display fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Hydrogen-peroxide cassette | Consumable | $$ | Easy |
| Vacuum pump / oil & filter kit | Periodic replacement | $$$ | Difficult |
| Chamber door seal | Periodic replacement | $$ | Moderate |
| Vaporizer plate | Periodic replacement | $$ | Fairly easy |
| Air / exhaust filter | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.