• User friendly software simplifies set-up and operation • Large & easily readable LCD display • Wide range of infusion rate • Over 40 types of compatible syringes (10-60 ml) • Drug dose calculation • Automatic & manual bolus capability • Automatic reduction of residual bolus after occlusion release • Easy modification of infusion parameters in running course • Comprehensive system of audible and visual alarms & warnings • Drug library with user selectable drug names • RS 232C interface • Long battery life • Data lock with a software-controlled passkey system • Event log (up to 2000 events recorded in real time): on-screen display or as XML file • 4 modes of operation: flow rate, flow rate and volume, flow rate and time, volume over time
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.