Arjohuntleigh AB · Model 526000-01
DVT Prevention Compression Pump - ACS900 PUMP US STD TUBE SET.
Intended use: A system used for the noninvasive treatment and prevention of venous disorders and sequelae such as oedema, deep vein thrombosis (DVT) and pulmonary embolism (PE). It functions by intermittently applying pressure to a patient's arms or legs to promote venous blood flow at timed intervals. This is achieved through the cyclical inflation and deflation of a single-chambered stocking, sleeve, or suit applied to the extremities. The system typically consists of a compression pump, control and timing mechanism, tubing, and pneumatic stocking(s) or suit.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Sleeve does not fill, low-pressure or garment-fault alarm, no compression cycle.
Root cause: Garment or tubing disconnected, kinked or leaking, wrong garment selected, or a pump / valve fault.
Symptoms: Pump audible but the garment does not alternate or hold.
Root cause: Rotary / solenoid valve or the manifold in the pump.
Symptoms: Delivered pressure does not match the set / rated value.
Root cause: Pressure transducer drift or a control-board fault.
Symptoms: Dead unit, or a persistent alarm.
Root cause: Power supply, fuse, or a control fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air pump | Major service part | $$ | Moderate |
| Solenoid / rotary valve assembly | Periodic replacement | $$ | Moderate |
| Pressure transducer | Periodic replacement | $$ | Moderate |
| Tubing set / connectors | Frequent wear item | $ | Easy |
| Compression garments / sleeves | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio