ARARCADIA GROUP · Model UA-767F-CP
Digital Blood Pressure Monitor.
Intended use: An electrically-powered device designed to noninvasively measure blood pressure using a self-contained software program that regulates automatic arm-cuff inflation and measurement cycles. It typically displays current heart rate and mean arterial pressure in addition to systolic and diastolic blood pressures; it may have memory to store blood pressure values and may sound an alarm if blood pressure exceeds pre-set limits. This device is not designed to be portable and is typically used bedside.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Times out, cuff will not inflate or hold pressure, readings implausible or fail against a reference.
Root cause: Internal pump or valve, a cuff / hose leak, or pressure-transducer / calibration drift.
Symptoms: Slow inflation, air leak, will not hold, artefact-heavy readings.
Root cause: Split cuff bladder, a cracked hose, or a loose connector.
Symptoms: Frequent retries or error on moving patients (common on ABPM).
Root cause: Algorithm limits plus a marginal cuff fit or transducer noise.
Symptoms: Short run time, will not power on or hold a 24 h ABPM study.
Root cause: Aged battery.
Symptoms: Study will not download, timestamps wrong.
Root cause: Download cable / dock or the internal clock battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Cuff and hose set | Frequent wear item | $ | Easy |
| Internal NIBP pump | Major service part | $$ | Moderate |
| NIBP valve block | Major service part | $$ | Moderate |
| Battery | Frequent wear item | $ | Easy |
| Pressure transducer / main board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.