ALPINION Medical Systems · Model E-CUBE 12
Storage: Combination of SSD 120GB for faster processing and HDD 1TB for saving large amount of patient data.3 Step Gel warmer for better patient comfort. Crystal Signature™ and Power View™ Transducer for better penetration and resolution. High density linear and High density Endocavity volume transducer for better resolution. Single crystal high density Convex transducer (1-6MHZ) for better penetration and resolution. Articulating monitor arm and Control panel up/down for height adjustment. Single crystal high density volume transducer (1-6MHZ) for reconstructing high quality 3D/4D images. Volume Master™- ALPINION’s intuitive 3D/4D software package in Convex and endocavity transducers offers fast rendering speed and outstanding rendering quality for diagnosis. Advanced 3D/4D features like Anyslice, Mslice and Volume Analysis for better 3D/4D diagnosis. Live HQ- Premium 3D/4D feature to render a realistic view of fetus. Advanced Cardiac features like stress echo, Cube strain, Auto EF, TDI. Other advanced features like Contrast Imaging, STIC imaging, Strain Elastography, Auto NT, Auto IMT, Needle vision Plus available. Touch Clip feature helps you to compare two different images for better diagnosis.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.