Aitecs · Model 12S AUTO
The AITECS 12 S AUTO has all the features of the 12 S Pro. It's the only syringe pump in the world capable of automatically increasing the flow rate at specified intervals during infusion. For example when used for Dobutamine infusions, the 12 S AUTO will auto increment the flow rate without user intervention. In addition, the 12 S AUTO has a drug library that is fully user programmable. These drug profiles enable the staff to consistently deliver medications as directed by the physician. Each 12 S AUTO is shipped with custom drug profiles made-to-order with your specific protocols. This unique drug library feature eliminates many of the steps required to prepare the pump for infusion, saving time and effort, which contributes to better patient care. AITECS Multi-Purpose Syringe Infusion System with Preset drug profiles and custom protocol programing software.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →