Air Liquide Healthcare · Model EOVE-150
Ventilator Based on an innovative blower design with fast response time down to 80ms, EOVE-150 ventilator offers best in class ventilation features for pediatric and adult patients with smart sensitive triggers and a wide range of tidal volume from 30 to 2500ml. Combining single or double branch circuit with tidal volume down to 30 ml and respiratory rate up to 80 bpm, EOVE-150 ventilators offers great performance to treat newborn patient.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable, replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.