Air Liquide Healthcare · Model CPX Care
The integrated medical air compressor. Medical air compressors provide a source of clean and dry compressed air. They can be used as main air source or as back-up in case of failure of the primary source.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low O2 purity, alarm.
Root cause: Zeolite sieve exhaustion.
Symptoms: Low flow, overheat.
Root cause: Worn compressor.
Symptoms: Low vacuum, no suction.
Root cause: Diaphragm/motor wear.
Symptoms: No humidity, overheat.
Root cause: Heater plate failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on concentrators.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Humidifier heating plate | Periodic replacement | $$ | Fairly easy |
| Sieve bed (zeolite) Common R&R on concentrators. | Periodic replacement | $$$ | Moderate |
| Compressor | Periodic replacement | $$$ | Difficult |
| Suction diaphragm / pump | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →