ACCUTRON, INC. · Model 35013
A device designed to hold and secure a single medical gas cylinder of compressed medical gas which is to be used in medical treatment/therapy. The usual gases are, [e.g., oxygen (O2), air, or nitrous oxide (N2O)]. It is used to secure the cylinder both during transport and typically whilst the gas is being used. The device is typically constructed of metal and may be mounted to a medical device service rail system, a bed under transport, or other medical devices requiring medical gas to be present. It is configured to meet the size of the cylinder, typically those of small dimensions.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Hiss when the cylinder is opened, cylinder empties quickly, frost on fittings.
Root cause: Missing or worn yoke washer / bodok seal, a loose gauge, or a diaphragm leak.
Symptoms: Flow does not match the setting, downstream device under- or over-supplied.
Root cause: Regulator diaphragm / spring drift, or a blocked inlet filter.
Symptoms: Wrong cylinder can be fitted, or the correct one will not seat.
Root cause: Damaged pin-index pins or the yoke.
Symptoms: Contents gauge reads high / low or is cracked.
Root cause: Worn or damaged Bourdon-tube gauge.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Yoke washer / bodok seals | Frequent wear item | $ | Easy |
| Inlet filter / sinter | Periodic replacement | $ | Fairly easy |
| Regulator diaphragm / spring kit | Major service part | $$ | Moderate |
| Contents gauge | Major service part | $ | Fairly easy |
| Pin-index pins | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.