Accusync Medical Research Corporation · Model 7200-5P
ACCUSYNC 72 5 LEAD CARDIAC TRIGGER MONITOR WITH PRINTER OPTION.
Intended use: A small unit dedicated to the measurement of cardiac output (the volume of blood pumped by the heart per minute) obtained through a pressure transducer/catheter introduced into a blood vessel. It is designed to operate as part of a patient monitoring system enhancing the function of this system (the parent device). The module automatically plugs into the parent device when the user places it into a standardized slot in the parent device or a connected module rack. The parent device operates as a mainframe computer displaying the cardiac output parameter/information measured and provided by this module.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Downstream device (IABP, imaging, lithotripter) not gating, missed or double triggers, noisy trigger.
Root cause: Poor electrode contact / skin prep, a worn 5-lead ECG cable, EMI, or a trigger-detection threshold set wrong.
Symptoms: Baseline wander, muscle noise, flat lead, "lead off".
Root cause: Dried or lifting electrodes, a damaged patient cable, or an input-amplifier fault.
Symptoms: ECG shows on the monitor but the trigger pulse does not reach the gated device.
Root cause: Trigger-output cable or connector, or a gating-output fault.
Symptoms: No or faint strip, paper jam.
Root cause: Thermal print head or paper handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| 5-lead ECG patient cable | Frequent wear item | $$ | Easy |
| ECG electrodes | Consumable | $ | Easy |
| Trigger-output cable | Periodic replacement | $$ | Easy |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Input amplifier board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio