Accuray Incorporated · Model 1056593
Software, Version 1.0.0.3, Treatment Delivery Console (TDC), Radixact (Operating Software for Radixact® Treatment Delivery System)
Intended use: An assembly of devices designed to produce high-energy electrons to yield high-energy x-rays used for therapeutic applications, primarily to treat cancer. It provides a uniform radiation field of uniform intensity and predictable energy level, in a beam with well defined dimensions. The principle of operation is linear acceleration of electrons by means of electromagnetic (EM) microwaves. The output is used as a treatment beam on the patient, either directly or after passing a suitable target. The system typically includes signal analysis and display devices, patient and device supports, and software.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Uncontrolled couch Z-axis movement (descent)
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio