Abiomed, Inc. · Model Impella 2.5 AIC Pump Set
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Intended use: A sterile tubular device with a built-in electrically-powered axial flow pump (e.g., a micro motor with an impeller) at the distal end, designed to provide circulatory assistance to the heart by pumping blood during heart failure. It is inserted through the lumen of the aorta and into the left ventricle, surgically or percutaneously via an artery, and connected to an external control unit; it moves blood from the left ventricle to the aorta across the aortic valve. It is typically used emergently post acute myocardial infarction (AMI), electively for high-risk cardiac catheterization, for open chest surgery, and during transport between hospitals. This is a single-use device.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
Pump not detected as connected to controller due to software issue.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.