Abiomed · Model Breethe OXY-1
The OXY-1 System is designed with an advanced gas exchange technology, including an innovative oxygen concentrator that provides full patient support with reduced O2 demand.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flow below target, flow alarms, pump speed / flow mismatch.
Root cause: Cannula position or kink, low circuit volume / venous collapse, an air lock, or a pump-head / drive fault.
Symptoms: Falling post-oxygenator O2, rising CO2, visible plasma leak or clot in the oxygenator.
Root cause: Oxygenator clotting or plasma breakthrough, or a sweep-gas supply / blender fault.
Symptoms: Bubble-detector alarm, pump stops.
Root cause: Air entrained at a connection or via the venous line, or a detector fault.
Symptoms: Short battery run time during transport, power alarm.
Root cause: Aged battery or charger fault.
Symptoms: Pre-/post-pump or oxygenator pressures implausible.
Root cause: Pressure transducer or its connection.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-use circuit component.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable oxygenator / pump-head module Single-use circuit component. | Consumable | $$$$ | Easy |
| Battery pack | Periodic replacement | $$$ | Fairly easy |
| Sweep-gas line / blender components | Periodic replacement | $$ | Moderate |
| Pressure transducers | Periodic replacement | $$ | Moderate |
| Bubble / air detector | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio