A & D Medical · Model One Step Plus Memory UA-767
Provides simultaneous readout of systolic/diastolic pressure and pulse rate. Calculates and displays average blood pressure readings. Alerts the user of the presence of an irregular heart beat during reading. Available in small, medium or large cuff sizes (cuffs can also be purchased separately). Clinically validated for accuracy. One button operation. 30 reading memory. Irregular Heartbeat feature. Large digital display. Fast measurement. Latex free.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Times out, cuff will not inflate or hold pressure, readings implausible or fail against a reference.
Root cause: Internal pump or valve, a cuff / hose leak, or pressure-transducer / calibration drift.
Symptoms: Slow inflation, air leak, will not hold, artefact-heavy readings.
Root cause: Split cuff bladder, a cracked hose, or a loose connector.
Symptoms: Frequent retries or error on moving patients (common on ABPM).
Root cause: Algorithm limits plus a marginal cuff fit or transducer noise.
Symptoms: Short run time, will not power on or hold a 24 h ABPM study.
Root cause: Aged battery.
Symptoms: Study will not download, timestamps wrong.
Root cause: Download cable / dock or the internal clock battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Cuff and hose set | Frequent wear item | $ | Easy |
| Internal NIBP pump | Major service part | $$ | Moderate |
| NIBP valve block | Major service part | $$ | Moderate |
| Battery | Frequent wear item | $ | Easy |
| Pressure transducer / main board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.