Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMINISTRATION SET, 138 IN.; Catalog Number: 490491.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0686-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/CARESITE EXT SET; Catalog Number: 490407.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0687-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/ B/C VALVE,103 IN.; Catalog Number: 470179.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0688-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, QUAD MANIFOLD,120 IN.; Catalog Number: 490340.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0689-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET,3 CARESITE LADS, 137 IN; Catalog Number: 490303.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0690-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 2 CARESITES, 125 IN.; Catalog Number: 490408.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0691-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. 130 IN. IV ADMIN SET W/ 9 IN. EXT SET; Catalog Number: 490412.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0692-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML 115 in.; Catalog Number: 490354.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0693-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 124 IN.; Catalog Number: 490451.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0694-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 15DR HF 4-WAY STPCK 135 IN.; Catalog Number: 490193.
2. IV AD SET, 2 CARESITE, SAFEPORT, 134 IN.; Catalog Number: 490326.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0695-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/ 15 DROPS/ML, 118 IN.; Catalog Number: 490437.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0696-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 128 IN.; Catalog Number: 480231.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0697-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 3 CARESITE LADS, 131 IN.; Catalog Number: 490274.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0698-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 103 IN.; Catalog Number: 490239.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0699-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. NF1505, IV ADMIN SET W/2 H.F. STOPCOCKS; Catalog Number: 490088.
2. IV ADMIN SET W/2 SAFELINE & HF STOPCOCK; Catalog Number: 490107.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0700-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. NF1500UP, IV ADMINISTRATION SET; Catalog Number: 490070.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0701-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 15DR, CARESITE LAD, 130 IN; Catalog Number: 477002.
2. IV ADMIN SET 85 IN.; Catalog Number: 477017.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0702-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV AD.SET W/15 DR 128 IN, EXT SET 9 IN; Catalog Number: 490301.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0703-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 129 IN, 3 CARESITE LADS; Catalog Number: 490225.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0704-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. AESTHESIA IV SET, 127 in.; Catalog Number: 490551.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0705-2026