Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 125 IN.; Catalog Number: 490487.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0666-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS, 162 IN.; Catalog Number: 490484.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0667-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 108 IN.; Catalog Number: 490504.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0668-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 15 DROP W/ CARESITE EXT SET; Catalog Number: 490528.
2. 111 IN. IV ADMIN SET W/ 12 IN. Y-EXT SET; Catalog Number: 490550.
3. IV ADMIN SET, 125 IN. w/ EXT SET 9 IN.; Catalog Number: 490605.
4. IV SET 15 DR 3CARESITE 2 ULTRAPORT; Catalog Number: 490616.
5. IV ADMIN SET, 2 CARSITE LADS; Catalog Number: 490654.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0669-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 15 DR W/2 SAFEDAY; Catalog Number: 352636.
2. IV ADMIN SET 15 DR W/2 SAFEDAY 84 IN.; Catalog Number: 352639.
3. IV ADMIN SET 15 DROP W/3 SAFEDAY; Catalog Number: 352640.
4. IV ADMIN 15D W/2 SAFEDAY & 0.2 FILTER; Catalog Number: 352643.
5. IV ADMIN SET, 105 IN.; Catalog Number: 477003.
6. IV ADMIN SET 15D, 4 CARESTIE LAD, 145 IN; Catalog Number: 477006.
7. IV ADMIN SET, 121 IN.; Catalog Number: 477008.
8. IV ADMIN SET 15D, CARESTIE LAD, 90 IN; Catalog Number: 477012.
9. IV ADMIN SET, 2 CARESTIE LAD, 15D, 117IN; Catalog Number: 477019.
10. IV ADMIN SET, 15D, 5CARESITE LAD, 136 IN; Catalog Number: 477021.
11. IV ADMIN SET 114 IN.; Catalog Number: 477022.
12. IV ADMIN SET 5 CARESITE LADS, 126 IN.; Catalog Number: 477023.
13. IV SET 15DR 3 SAFEDAY, 106 IN.; Catalog Number: 477053.
14. IV ADMIN SET, 15DR, CARESITE LAD, 121 IN; Catalog Number: 490513.
15. IV ADMIN SET, 15D, 120 IN.; Catalog Number: 490521.
16. IV ADMIN SET 15D 2 CARESITE LADS, 112IN.; Catalog Number: 490527.
17. IV ADMIN SET 15 DRP CARESITE SAFEPORT; Catalog Number: 490529.
18. IV ADMIN SET, 15D, 139 IN.; Catalog Number: 490535.
19. IV ADMIN SET 15D, CARESITE LAD, 122IN.; Catalog Number: 490543.
20. IV ADMIN SET 15D 3 CARESITE LADS, 150IN.; Catalog Number: 490546.
21. IV ADMIN SET, 125 IN.; Catalog Number: 490547.
22. IV ADMIN SET 15D, 5 CARESITE INJ, 150in.; Catalog Number: 490552.
23. IV ADMIN SET, 120IN.; Catalog Number: 490553.
24. IV ADMIN SETW/ B/C VLV 3 CARESITE 134IN.; Catalog Number: 490554.
25. IV ADMIN SET 15DR w/3 B/C VLV, 124 IN.; Catalog Number: 490555.
26. IV ADMIN SET 15D w/B/C VLV, 139 IN.; Catalog Number: 490558.
27. IV ADMINISTRATION SET, 117 IN.; Catalog Number: 490559.
28. IV ADMINISTRATION SET, 108 IN.; Catalog Number: 490560.
29. IV ADMINISTRATION SET, 134 IN.; Catalog Number: 490561.
30. IV ADMIN SET 15D w/B/C VLV, 140 IN.; Catalog Number: 490562.
31. IV ADMIN SET, 111 IN.; Catalog Number: 490563.
32. IV ADMIN SET 15D 2 CARESITE LADS, 122 IN; Catalog Number: 490569.
33. IV ADMIN SET 15D STOPCOCK, DUAL MANIFOLD; Catalog Number: 490573.
34. IV ADMIN SET 15D w/B/C VLV, 110 IN.; Catalog Number: 490574.
35. IV ADMIN 15D 5 CARESITE LADS STPCK 120in; Catalog Number: 490575.
36. IV ADMIN SET 15D 5 CARESITE LADS, 140 IN; Catalog Number: 490578.
37. IV ADMIN SET W/15D, 4 CARESITE LADS; Catalog Number: 490581.
38. IV ADMIN 15D CARESITE LADS, STPCK 134 IN; Catalog Number: 490582.
39. IV ADMIN SET, 15D, 108 IN.; Catalog Number: 490583.
40. IV ADMIN SET, 10 DROP, 130 IN.; Catalog Number: 490588.
41. IV ADMIN SET, 132 IN.; Catalog Number: 490589.
42. IV ADMIN SET, 15DR, 105 IN.; Catalog Number: 490600.
43. IV ADMIN SET,3 CARESITE,3 HF SPCK, 129IN; Catalog Number: 490607.
44. IV SET 15DR 3CARESITE 2STOPCOCK, 126IN; Catalog Number: 490608.
45. IV ADMIN SET 15D 2 CARESITE LADS, 138 IN; Catalog Number: 490615.
46. IV SET 15DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490618.
47. IV SET 60DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490619.
48. IV ADMIN SET 15 DRPS 4W STOPCOCK W/ C&S; Catalog Number: 490644.
49. IV ADMIN SET 15 DRPS ULTRAP. ZERO 90.6IN; Catalog Number: 490657.
50. IV ADMIN SET 15 DRPS STOPCOCK MALE 117IN; Catalog Number: 490658.
Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/ 15 DROPS/ML, 99 IN; Catalog Number: 490511.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0671-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 15 DROP, 132 IN; Catalog Number: 490519.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0672-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 15DR, CARESITE LAD, 116 IN; Catalog Number: 490522.
2. IV ADMIN SET 115, IN. W. 9 IN EXT SET; Catalog Number: 490568.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0673-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMINISTRATION SET WITH 15 DROPS/ML; Catalog Number: 490507.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0674-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Catalog Number: 490549; Primary UDI-DI: 04046955270323; Unit of Dose UDI-DI: 04046955270316; Expiration Date: Earliest Exp of Comp or 36mths.
2. Catalog Number: 490565; Primary UDI-DI: 04046955275601; Unit of Dose UDI-DI: 04046955275595; Expiration Date: Earliest Exp of Comp or 36mths.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0675-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 133 IN.; Catalog Number: 490245.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0676-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. 15DR 2SAFELINE 1CARESITE 1STOPCOCK 131IN; Catalog Number: 490612.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0677-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 15D, 2 CARESITES, 153 IN.; Catalog Number: 490542.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0678-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. PB1500 15 DROP IV SET W/2 STOPCOCKS; Catalog Number: 490025.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Accessories, CatheterDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0679-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. NF1503 15 DR UNIV SPIKE, INJ SITE, BCV; Catalog Number: 490062.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Accessories, CatheterDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0680-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/10 DROPS, 138 IN.; Catalog Number: 490426.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Accessories, CatheterDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0681-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 126 IN.; Catalog Number: 490488.
2. IV ADMIN SET W/15 DROPS/ML, 132 IN.; Catalog Number: 490492.
3. IV ADMIN SET W/15 DROPS/ML, 139 IN.; Catalog Number: 490503.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0682-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADM 15 DROP W/3 CARESITE N/C DUAL; Catalog Number: 354211.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0683-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADM 15D 2 CARESITE, SAFEPORT MANIFLD; Catalog Number: 354208.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0684-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMN SET W/15 DROP, ULTRAPORT,134 IN.; Catalog Number: 490312.
2. IV ADMIN SET W/ 15 DROPS/ML, 120 IN.; Catalog Number: 490432.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0685-2026