Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Primary Gravity Filtered Set with 2 CARESITE Injection Sites; IV ADM 15 DROP W/2 CARESITE FILTER; Catalog Number: 354207.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0626-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. AS152V2DLL; Catalog Number: 473309.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0627-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. ET08US EXTENSION SET W/ULTRA Y SITE; Catalog Number: 474912.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0628-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. ANES SET W/DUAL FEM.LL SAFEPORT MANIFOLD; Catalog Number: 456520.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0629-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. EXTENSION SET, 18 IN; Catalog Number: 490448.
2. EXTENSION SET, 52 IN.; Catalog Number: 490465.
3. EXTENSION SET 49 IN.; Catalog Number: 490486.
4. EXTENSION SET, 12 CARESTITE LADS; Catalog Number: 490577.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0630-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXTENSION SET, 25 IN.; Catalog Number: 490629.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0631-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXT SET W/ B/C VLV, ULTRAPORT, 57 IN.; Catalog Number: 490202.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0632-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXTENSION SET W/ B/C VALVE 53 in.; Catalog Number: 490358.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0633-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXTENSION SET W/ ULTRAPORT, 48 IN.; Catalog Number: 490291.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0634-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXT SET ULTRAPORT STOPCOCK, 37IN.; Catalog Number: 490234.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0635-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXT SET W/ 6 B/C VALVES, 17 IN; Catalog Number: 490238.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0636-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. ANESTHESIA SET W/ SAFEPORT MANIFOLD; Catalog Number: 456521.
2. ANES SET W/TRIPLE LL SAFEPORT MANIFOLD; Catalog Number: 456522.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0637-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Caresite Extension Set_Export; Catalog Number: 470183.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0638-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. EXT/2INJ/FLBL/FML LRLK/CV 34; Catalog Number: V5405.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0639-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032.
2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033.
3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034.
4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine; Catalog Number: 362043.
5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050.
6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420.
7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431.
8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432.
9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010.
10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019.
11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030.
12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230.
13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419.
14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420.
15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421.
16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422.
17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423.
18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424.
19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430.
20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433.
21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN.
CLOSE"; Catalog Number: 480254.
22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255.
23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256.
24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258.
25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263.
26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100.
27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.
Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/ 10 DROPS/ML, 134 in.; Catalog Number: 490357.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0641-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. IV ADMIN SET, 10 DROPS/ML, 144 IN; Catalog Number: 490385.
2. IV ADMIN SET, 3 CARESITES, 120 IN.; Catalog Number: 490401.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0642-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. NF1270A SAFELINE ANESTHESIA IV SET; Anesthesia IV Administration Set with 3 CARESITE Injection Sites and Manifold; Catalog Number: NF1270A.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0643-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. US1295 ULTRASITE IV ADMIN SET 60DROP; Catalog Number: 352394.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0644-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV SET 15DR 5CARESITE 1 ULRAPORT, 141 IN; Catalog Number: 490067.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0645-2026