Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 108 IN; Catalog Number: 490189.
2. IV ADMIN SET 137 IN.; Catalog Number: 490226.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0606-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 132 IN; Catalog Number: 490469.
2. IV ADMIN SET, 60D 2 CARESITE VLV, 132 IN; Catalog Number: 490470.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0607-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV AD. SET,15 DROPS/ML, W/EXT SET; Catalog Number: 490467.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Stopcock, I.V. SetDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0608-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/15 DROPS/ML, 134 IN.; Catalog Number: 490477.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0609-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 15D, 144 IN.; Catalog Number: 490472.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0610-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 15 DROPS/ML 125 IN; Catalog Number: 490452.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0611-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 15D 2 CARESITE LAD, 126 IN; Catalog Number: 490444.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0612-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET, 2-HIGH FLOW STPCKS, 124 IN; Catalog Number: 490476.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0613-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET 117 IN.; Catalog Number: 490468.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0614-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
B.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. IV ADMIN SET W/ 60 DROPS/ML,141 in.; Catalog Number: 490349.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0615-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Y-TYPE BLOOD SET 124 IN.; Catalog Number: 490196.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0616-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Y-TYPE BLOOD SET W/ 170 ?M FILTER 120 IN; Catalog Number: 490355.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0617-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Y-TYPE BLOOD SET, 3 CARSITE LADS, 146 IN; Catalog Number: 490530.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0618-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Y-TYPE BLOOD SET, HAND PUMP, 146 IN.; Catalog Number: 490438.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0619-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. Y-TYPE BLOOD SET W/ 170M FILTER; Catalog Number: 480233.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0620-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. BURETTE SET, 4 CARESITE LADS, STPCK 125"; Catalog Number: 490252.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0621-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number:
1. BURETTE SET W/ 2 CARESITE, 134 IN.; Catalog Number: 490159.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0622-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
CARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. CARESAFE IV ADMIN SET w/AIRSTOP, 104 IN.; Catalog Number: 354302.
2. CARESAFE ADMIN SET w/15um FILTER, 104 IN; Catalog Number: 354303.
3. CARESAFE IV ADMIN SET w/AIRSTOP, 114 IN.; Catalog Number: 354304.
4. CARESAFE ADMIN SET w/15um FILTER, 114 IN; Catalog Number: 354305.
5. CARESAFE IV ADMIN SET w/AIRSTOP, 110 IN.; Catalog Number: 354306.
6. CARESAFE ADMIN SET w/15um FILTER, 110 IN; Catalog Number: 354307.
7. CARESAFE" IV Administration Set with AirStop Filter, Not made with PVC or DEHP, 2 CARESITE Injection Sites, 2 ULTRAPORT High-FlowFour-Way Stopcocks; CARESAFE ANESTH SET W/AIRSTOP, 134 IN.; Catalog Number: 354308.
8. CARESAFE ANESTH SET w/15um FILTER,134IN.; Catalog Number: 354310.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0623-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. Primary Gravity IV Set with 2 CARESITE Injection Sites; IV ADMIN SET 60 DROP W/2 CARESITE; Catalog Number: 354204.
2. Primary Gravity IV Set with 3 CARESITE Injection Sites; IV ADMIN SET 15 DROP W/3 CARESITE; Catalog Number: 354205.
3. Primary Gravity IV Set with 3 CARESITE Injection Sites; IV ADMIN SET 60 DROP W/3 CARESITE; Catalog Number: 354206.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0624-2026 Open, ClassifiedPosted Nov 26, 2025·Initiated Oct 29, 2025
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers:
1. Primary Gravity IV Set with 2 CARESITE Injection Sites; IV ADMIN SET 15 DROP W/2 CARESITE; Catalog Number: 354202.
2. Primary Gravity IV Set with 2 CARESITE Injection Sites; IV ADM 15DROP W/2 CARESITE SLIDECLMP; Catalog Number: 354203.
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root cause: Process control
Recalling firm: B Braun Medical IncType: Set, Administration, IntravascularDistribution: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.Recall #Z-0625-2026