ZOLL Medical · Model E Series
With ZOLL's Real CPR Help®, chest compressions and rate and depth is measured in real time, and visual and optional audible feedback are provided. All CPR data can be recorded and reviewed using RescueNet Code Review software. Unique to ZOLL, our See-Thru CPR® allows CPR artifact to be filtered, allowing you to see organized rhythms without pausing compressions. New innovative roll cage design and shape. EasyRead Tri-Mode Display™ screen for use in both pitch darkness or direct sunlight. A new Rapid Cable Deployment™ System for faster deployment. Built-in GPS clock, allowing users to synchronize dispatch, defibrillator, and intervention call times, improving overall data accuracy. Compatible with ZOLL’s sealed lead acid 4410 Smart Battery™ as well rechargeable lithium ion. Integrates optional Bluetooth® functionality to provide data transmission capability to a variety of destinations. Uses ZOLL’s proprietary Rectilinear Biphasic™ waveform, the only biphasic waveform cleared by the FDA to claim superiority to monophasic waveforms for the conversion of ventricular fibrillation in high impedance patients.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Delivered energy outside tolerance on the analyser, will not charge, or charges slowly.
Root cause: HV capacitor / charging-circuit ageing, or an output-relay fault.
Symptoms: Short standby, readiness indicator shows service, unit fails the self-test on battery.
Root cause: Aged battery pack or a charging fault.
Symptoms: "Check pads", "connect electrodes", no ECG through pads, poor paceing capture.
Root cause: Worn therapy cable, expired pads, or a paddle / connector fault.
Symptoms: A monitoring parameter drops or is implausible.
Root cause: Patient cable / sensor or the parameter module.
Symptoms: No printout, event data will not transfer.
Root cause: Thermal recorder or the data / Wi-Fi interface.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Easy |
| Therapy / defib cable | Periodic replacement | $$ | Fairly easy |
| Multifunction pads | Frequent wear item | $ | Easy |
| HV / charging board | Major service part | $$$$ | Difficult |
| Thermal recorder / print head | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.