ZOLL Medical · Model AED Pro
Superior design and rubber over-molds are built to withstand harsh environments. Real CPR Help for rate and depth of chest compressions during CPR using ZOLL's unique one-piece CPR-D-padz®. See-Thru CPR allows a rescuer to view a patient's underlying ECG rhythm while performing CPR. Exceptionally resistant to dust and water jets. Only AED in the industry to withstand a 1.5-meter drop test, making it suitable to withstand the rough treatment typically found in an emergency medical environment. Built-in internal memory and external USB memory technology allows expanded data storage capacity, and maximum flexibility for data management and transfer. Electrodes and battery compatibility with ZOLL M Series®, E Series® and R Series® professional defibrillators. ECG Monitoring with 3-lead cable and high-quality display. Unrivaled battery options (long-life disposable or rechargeable) that meet your specific needs. Easy unit configuration through ZOLL's administration software.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Charge fails or sluggish, low-energy delivery.
Root cause: Capacitor aging / leakage.
Symptoms: No shock delivered, pad error.
Root cause: Worn paddles or expired pads.
Symptoms: Won't charge or discharges unexpectedly.
Root cause: Relay/contactor wear.
Symptoms: Leakage alarm, self-test fail.
Root cause: Failed isolation components.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Critical, keep charged spares.
Expiration-dated consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack Critical, keep charged spares. | Periodic replacement | $$$ | Fairly easy |
| HV energy storage capacitor | Major service part | $$$$ | Expert-level |
| Defibrillation electrodes / pads Expiration-dated consumable. | Consumable | $$ | Easy |
| Charge relay / contactor | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →