Heart defibrillator has a built in Text Display with Voice Prompts FSA Eligible Analyzes heart rhythm and informs the rescuer if the rhythm is shockable or non-shockable. Delivers defibrillation treatment to victims of cardiac arrest who exhibit shockable ECG rhythms. Sequenced LED array gives fast communication of specific rescuer actions needed at the correct time. Voice prompts and graphics guide the rescuer regarding what to do and when to do it during a cardiac emergency - such as calling for help or giving CPR to the patient. Monitors the depth and rate of chest compressions during CPR and provides voice prompts Notifies user of good compressions that are at least 2 inches deep and at a rate of 100 per minute Notifies users if compressions are inadequate and when to push harder and/or faster Meets the FDA's guidelines for requiring that chest compressions be at least 2 inches deep Audible beeps encourage rescuers to provide CPR compressions at 100 CPM (requires Zoll CPR-D Padz) Performs periodic self test to ensure its continual readiness Unit cover functions as a Passive Airway Support System (PASS). Resists dust and water.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Delivered energy outside tolerance on the analyser, will not charge, or charges slowly.
Root cause: HV capacitor / charging-circuit ageing, or an output-relay fault.
Symptoms: Short standby, readiness indicator shows service, unit fails the self-test on battery.
Root cause: Aged battery pack or a charging fault.
Symptoms: "Check pads", "connect electrodes", no ECG through pads, poor paceing capture.
Root cause: Worn therapy cable, expired pads, or a paddle / connector fault.
Symptoms: A monitoring parameter drops or is implausible.
Root cause: Patient cable / sensor or the parameter module.
Symptoms: No printout, event data will not transfer.
Root cause: Thermal recorder or the data / Wi-Fi interface.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Easy |
| Therapy / defib cable | Periodic replacement | $$ | Fairly easy |
| Multifunction pads | Frequent wear item | $ | Easy |
| HV / charging board | Major service part | $$$$ | Difficult |
| Thermal recorder / print head | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.