Ziehm Imaging · Model Vista Plus
Full Frame Technology does away with inadequate resolution, creating superb images without artefacts. The intelligent Ziehm Anatomical Programs such as Bone/Extremities, Pelvic/Spine and Thorax/Abdomen ensure superb image quality in each type of procedure. In addition "Metal" and "LPD" (Large Patient Diameter) image programs further enhance image quality in otherwise difficult imaging conditions. Intuitive user interface and logical user guidance on-screen make it easy to use the system’s comprehensive features with no need for extensive training of staff. Light-weight architecture and mechanical precision enable perfect clinical mobility and effortless positioning. The unique combined steering and breaking control lever permits easy and precise positioning around the operation table. Active Cooling allows for considerably prolonged fluoroscopy times. With the modular upgrade path of the product family, the standard model Ziehm Vista Plus can be upgraded with vascular functionality. Optional comprehensive access to the DICOM world (Ziehm NetPort).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No image on boost, exposure inhibited, kV/mA error, image too dark or bright.
Root cause: X-ray tube / monoblock fault, HV generator fault, or the AEC / dose-control loop.
Symptoms: Dim, distorted, vignetted or noisy image; dead lines (flat panel); S-distortion (II).
Root cause: Image-intensifier tube ageing / II fault, or a flat-panel detector fault.
Symptoms: Orbital, lateral or vertical motion binds, drifts, or a brake will not hold.
Root cause: Worn friction brakes, bearings, or the counterbalance.
Symptoms: No image on one or both monitors, image freeze, no last-image-hold.
Root cause: Monitor, the cart-to-stand cable, or the image-processing PC.
Symptoms: Fluoro inhibited after a run, tube-heat warning.
Root cause: Monoblock heat limit reached or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| X-ray tube / monoblock | Major service part | $$$$ | Expert-level |
| Image intensifier tube / camera | Major service part | $$$$ | Difficult |
| Flat-panel detector (if fitted) | Major service part | $$$$ | Difficult |
| C-arm friction-brake / bearing kit | Major service part | $$ | Moderate |
| Monitor / video cable | Major service part | $$ | Fairly easy |
| Cooling fan | Periodic replacement | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.