Yon Don · Model YD-F-285TR
Digital microprocessor based control system. 3-Level over-heating protection mechanism. Skin sensor failure detection (only available on 285TRSC). Power failure detection. Reminder audio alerts every 15 minutes if actual air/skin temperature does not reach within 1 degree range of the set temperature. Advanced programmable heating parameters for different ambient conditions. Touch keys for easy control settings. Water-proof control panel film to protect against spills or splashes. Large LED displays for easy readings on incubator status. Durable and transparent acrylic provides great visibility to infant. Convenient access ports reduce air temperature fluctuations. Double-wall minimizes temperature fluctuations (optional feature). Soft mattress provides comfort to the infant. Advanced air circulation system to provides best exchange and distribution of heat and air/oxygen. Spacious compartment provides room for oxygen hood (hood is optional). Convenience ports available to allow secure pathways for skin probes (only available on 285TRSC), oxygen tubes, and etc. Front door allows transporting infant in/out of the incubator with great ease. Tilt-enabled compartment provides maximum access to interior of the compartment.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Chamber / patient will not reach or hold the set temperature, temperature alarm, self-test / "Check" fault.
Root cause: Heater element / infrared rod, the air- or skin-temperature sensor, the fan (incubator), or a control-board fault.
Symptoms: Servo mode does not regulate to the skin probe, deviation alarm, probe-fault alarm.
Root cause: Damaged skin temperature probe or its connector, or the servo-control circuit.
Symptoms: No or quiet audible alarm, alarm lamp does not light on the self-test.
Root cause: Buzzer / speaker or the alarm board - a safety-critical failure.
Symptoms: Set humidity not reached, uneven chamber temperature, airflow alarm.
Root cause: Humidifier reservoir / heater, or a worn circulation fan / blocked air filter.
Symptoms: In-bed scale inaccurate, tilt mechanism stiff, port latches or iris sleeves worn.
Root cause: Load cell, tilt actuator, or worn access-port hardware.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air / skin temperature sensors | Frequent wear item | $$ | Fairly easy |
| Heater element / infrared rod | Major service part | $$ | Difficult |
| Circulation fan (incubator) | Major service part | $$ | Moderate |
| Humidifier heater / reservoir (incubator) | Periodic replacement | $$ | Moderate |
| Buzzer / alarm board | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.