Xoran Technologies · Model MiniCAT IQ
Sharper Imaging Simpler workflow Single Visit Diagnosis Socially distant CT imaging Lower Radiation Dose Scan and reconstruct in under a minute
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Tube won't fire, heat-capacity alarm.
Root cause: Filament wear, cooling failure.
Symptoms: Image noise, artifacts, low SNR.
Root cause: Flat-panel / sensor aging.
Symptoms: Exposure fails, kV/mA out of range error.
Root cause: HV tank or generator board fault.
Symptoms: Frozen console, acquisition timeout.
Root cause: Workstation or RAID degradation.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flat-panel detector | Major service part | $$$$ | Expert-level |
| Cooling pump & radiator | Periodic replacement | $$$ | Difficult |
| High-voltage generator | Major service part | $$$$ | Expert-level |
| Workstation RAID / GPU | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Out of tolerance kVp, Unexposed (black) frames, and Out of tolerance Dose.
A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
Misbranding; These devices do not bear the manufacturer's name or address.
The hard drive may crash, resulting in loss of data from patient CT scans.