Vyaire · Model Vyntus CPX
Gas analyzer technology improvements – 75ms response time and a unique “Tool-free“ O2 fuel cell change Proven Digital Volume Transducer (DVT) flow sensor technology provides accurate and reliable flow/volume measurements in the complete range of low-to-high flow Interfaces seamlessly with numerous ECGs for comprehensive CPET testing on a single curved wide screen, which we call Big Cinema. Automated volume calibration ensures consistency, saves time and hassle On-board pulse oximetry with finger, ear-clip and forehead sensors Powered by SentrySuite™ with cues and guidance during measurement and post-test workflow to help standardize evaluations and reduce time to results Smart tools automate pre-test setup and provide timed cues & guidance for events required during the test (e.g.: blood pressure, RPE signaling, exercise flow volume loops, etc.)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: 3 L syringe check outside tolerance, volume / flow drift over the day, biological control out of range.
Root cause: Flow-sensor contamination or wear (pneumotach / turbine / ultrasonic), a leak, or ambient (BTPS) correction not set.
Symptoms: Erratic traces, poor repeatability, back-extrapolated volume errors.
Root cause: Worn or wet flow sensor, a damaged filter seat, or a leak in the tubing / mouthpiece connection.
Symptoms: DLCO or FRC results implausible, gas-analyser calibration fails, drift.
Root cause: CO / CH4 / He / tracer analyser drift or an analyser fault, or a demand-valve / gas-supply issue.
Symptoms: Box pressure will not stabilise, thermal drift, door-seal leak.
Root cause: Door seal, the box pressure transducer, or thermal equilibration.
Symptoms: Test hangs, results will not save, print or export.
Root cause: Workstation PC, software, or the printer.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Calibrate / verify the syringe periodically.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor (pneumotach / turbine / ultrasonic) | Periodic replacement | $$ | Fairly easy |
| Calibration syringe (3 L) Calibrate / verify the syringe periodically. | Major service part | $$ | Easy |
| Gas analyser cell (DLCO / lung-volume systems) | Periodic replacement | $$$ | Moderate |
| Plethysmograph door seal / pressure transducer | Periodic replacement | $$ | Moderate |
| Sample lines / demand valve | Frequent wear item | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.