MGC Diagnostics · Model Platinum Elite
Eliminates warm-up or recalibration between patients. Contains no moving parts or electronics for cost-effective testing. Ease of use allows operator to focus on patient's effort. Comprehensive diagnostics, including airways resistance and thoracic gas volumes. Automated quality-assurance program notifies operator to ensure results comply with ATS/ERS acceptability criteria. Can be sterilized, used with a filter or discarded, making it extremely versatile. Electronic medical record interface using latest HL7 standards with either unidirectional or bidirectional capability. Flexible access allows physicians to view and interpret patient results from any location on the facility’s network.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Zeroing fails, cal factor outside 1.6-2.2, flow/volume values inaccurate.
Root cause: Air passing through the flow sensor, an obstruction in the sensor or umbilical, the sensor not dried after sterilisation, or cal factors drifted.
Symptoms: A test phase does not run, valve does not switch, WFA Tester shows no response on a channel.
Root cause: Failed solenoid valve or pump, or a driver on the WFA / control board.
Symptoms: DLCO analyse trace abnormal, gas analysis fails, drift over time.
Root cause: Gas chromatograph column degraded or contaminated, or an analyzer calibration issue.
Symptoms: Box leaks, pressure test fails, seal sticky or dry-rotted.
Root cause: Door seal drying out - needs silicone lubricant or cornstarch treatment, or replacement if damaged.
Symptoms: System power indicator not amber/green, LED1 on the Flow PCB not flashing ~1 Hz.
Root cause: Power supply or Flow PCB fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Wear/consumable item; keep spares and dry fully after sterilisation.
Do not clean with solvents; use only non-toxic silicone lubricant (MGC P/N 104185-001).
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| preVent flow sensor Wear/consumable item; keep spares and dry fully after sterilisation. | Frequent wear item | $$ | Fairly easy |
| Solenoid valve / pump | Major service part | $$ | Moderate |
| Gas chromatograph column | Periodic replacement | $$$ | Moderate |
| Body-box door seal Do not clean with solvents; use only non-toxic silicone lubricant (MGC P/N 104185-001). | Periodic replacement | $$ | Moderate |
| Flow PCB | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.