Verathon · Model GlideScope AVL
Ideal for difficult and unpredictable airways; convenient for routine airways Provides a consistently clear view, enabling quick intubation in unpredictable conditions Operational in seconds Optimized for demanding OR, ED, and ICU settings.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Black screen, snow, distorted or dim image, no image with a blade connected.
Root cause: Failed video baton/camera or its cable, a reusable blade at end of life, or optics contamination/fog.
Symptoms: Short run time on battery, unit will not power on, will not charge in the dock.
Root cause: Aged internal battery or a charging-circuit / dock fault.
Symptoms: Blade tip cracked or separated (subject of an FDA recall), poor fit on the baton.
Root cause: Fatigue/premature failure of the reusable GVL/AVL blade tip, or an out-of-validation blade combination.
Symptoms: Unresponsive buttons, image freeze, software behaviour not matching the manual.
Root cause: Monitor hardware fault or software out of date (contact Verathon for upgrades).
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Not user-repairable internally; replace as a unit.
Service by a qualified technician per Verathon.
Inspect the tip before every use; retire per the recall guidance.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Video baton / camera stick Not user-repairable internally; replace as a unit. | Periodic replacement | $$$ | Easy |
| Internal battery Service by a qualified technician per Verathon. | Periodic replacement | $$ | Fairly easy |
| Reusable video laryngoscope blade (GVL/AVL) Inspect the tip before every use; retire per the recall guidance. | Periodic replacement | $$ | Easy |
| Video monitor assembly | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Product labeled as GVL 2 Stat (blades), lot GC32825, was physically packaged with the incorrect size device (GVL 1 Stat (blades)).
Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.
Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio